Wednesday, July 29, 2009

FWHM in CONUSS

Total number of points which can be assigned for the width of EFG in CONUSS is limited. If too many points are assigned for the width, CONUSS crashes but does not specify what is wrong.

Dan Shim learn from Brent Grocholski

Ar loading

In order to prevent blow out, Ar loading should be done before or at least at noon. We don't know why. But we found there is much less chance of blow out in morning loadings. This is particularly true in Summer time when humidity is high.

Tuesday, May 26, 2009

Make Ubuntu+CONUSS faster

Make sure you have the following commands in Ubuntu.bat file.

-----Ubuntu.bat--------------
REM Start qemu on windows.
@ECHO OFF

net start kqemu

qemu -L . -kernel-kqemu -localtime -soundhw all -hda hda.img -m 512

Do not load Ar without humidity control

Cryogenic Ar loading should be done at humidity below 40%. Otherwise, sample chamber would expand during or after the loading

Dan Shim